GW Instek GPE-X323A Series, Multi-Channel Linear DC Power Supplies Introduction
The GW Instek GPE-X323A series represents the evolution of the esteemed GPE-X323 line, integrating advanced digital control, automatic tracking functions, and comprehensive protections. With a 4.3" LCD display and optimized heat dissipation design, this series is ideal for precision testing and extended-duration applications.
- 4 models: GPE-1326A (1 channel), GPE-2323A (2 channels), GPE-3323A (3 channels), GPE-4323A (4 channels)
- Independent isolated channels with automatic series/parallel functions
- Digital control with encoder switch for improved accuracy and durability
- High resolution: 10 mV/1 mA for setting and reading
- Integrated protections: OVP, OCP, OTP and panel lock function
- 4.3" monochrome LCD display for clear parameter visualization
Target Applications
- Scientific research and experimental testing
- Applications requiring low noise and stable voltage output
- Electronic component measurements
- 3C electronics product testing (Computer, Communication, Consumer)
Technological Innovations
- Encoder switch control: Replaces conventional VR knobs for greater accuracy and durability
- Advanced OVP protection: Individual settings for each channel for maximum flexibility
- Set View function: Check parameters without disabling output
- Panel Lock: Protection against accidental operations during extended tests
- Optimized heat dissipation: Improved thermal design for continuous operation
| Model |
CH1 |
CH2 |
CH3 |
CH4 |
Total Power |
| GPE-1326A |
32V / 6A |
- |
- |
- |
192W |
| GPE-2323A |
32V / 3A |
32V / 3A |
- |
- |
192W |
| GPE-3323A |
32V / 3A |
32V / 3A |
5V / 5A |
- |
217W |
| GPE-4323A |
32V / 3A |
32V / 3A |
5V / 1A |
15V / 1A |
212W |
Key Technical Specifications
- Line regulation: ≤0.01% + 3 mV
- Load regulation: ≤0.01% + 3 mV (≤3A) / ≤0.02% + 5 mV (>3A)
- Ripple and noise: ≤1 mVrms (5 Hz - 1 MHz)
- Recovery time: ≤100 μs (50% load change, min load 0.5A)
- Setting/reading resolution: 10 mV / 1 mA
- Accuracy: ±(0.1% of reading + 30 mV) voltage, ±(0.3% of reading + 6 mA) current
Automatic Series/Parallel Functions
- Series mode (CH1/CH2): 0-64V / 0-3A (doubles voltage)
- Parallel mode (CH1/CH2): 0-32V / 0-6A (doubles current)
- Tracking error: ≤0.1% + 10 mV of Master (no load)
- Automatic internal connection: CH1/CH2 without external wiring
Complete Protection System
- OVP (Over Voltage Protection): Range 1-36V (CH1/CH2), 1-6V (CH3), 1-16V (CH4)
- OCP (Over Current Protection): Programmable overcurrent protection
- OTP (Over Temperature Protection): Automatic thermal protection
- Reverse polarity protection: Integrated on all channels
- Isolation: 20 MΩ (terminals-chassis), 30 MΩ (chassis-AC)
Operating and Environmental Specifications
- Power supply: AC 100/120/220V ±10%, 230V +10%/-6%, 50/60 Hz
- Consumption: 550 VA / 420 W max
- Operating temperature: 0°C to 40°C, humidity ≤80%
- Dimensions: 210 × 155 × 306 mm (W×H×D)
- Weight: Approximately 7 kg
Standard Equipment
- Included: Power cord, component list
- European test leads:
- GPE-1326A: GTL-204A × 1, GTL-203A × 1
- GPE-2323A: GTL-204A × 2
- GPE-3323A: GTL-204A × 3
- GPE-4323A: GTL-204A × 2, GTL-203A × 2
Remote Control
- Remote control terminal: ON/OFF outputs via rear connector
- Pin configuration: Pin 7&8 (setting), Pin 9&10 (ON/OFF control)
- Status indication: Flashing ON/OFF icons during remote control